Medical Device Repair Cases

Quality SonoScape EG-500 Electronic Gastroscopy For Repair factory

SonoScape EG-500 Electronic Gastroscopy For Repair

SonoScape EG-500 Electronic Gastroscopy For Repair ► Brief Description Inspection Results: 1. No image display, CMOS water ingress fault; 2. 20% of the light bundle is broken; 3. There are bite marks on the insertion tube in multiple places; 4. The biopsy tube is worn out and not smooth; 5. The angle is seriously insufficient and tight, with wear on the snake bone, steel wire, and helix tube; 6. The light guide tube has dents; 7. The handle shell is cracked and has severe
Quality Olympus ShockPULSE-SE Lithotripsy Handpiece SPL-T For Repair factory

Olympus ShockPULSE-SE Lithotripsy Handpiece SPL-T For Repair

Olympus ShockPULSE-SE lithotripsy handpiece SPL-T for repair ► Brief Description Fault: 1.No ultrasonic energy output. Ultrasonic crystal is damaged. Solution: Replace the ultrasonic crystal. ►Problem & Solutoin Brand Olympus Model SPL-T Ultrasonic crystal Damaged Fault No ultrasonic energy output Solution Replace the ultrasonic crystal ► Applications The Olympus ShockPULSE-SE lithotripsy handpiece (SPL-T) is a key component of the Olympus ShockPULSE-SE Lithotripsy System,
Quality OLYMPUS OTV-S6H-1D Camera Head For Repair factory

OLYMPUS OTV-S6H-1D Camera Head For Repair

Olympus OTV-S6H-1D Camera Head For Repair ► Brief Description Fault: 1.Image yes. 2.Cable aging and interference Solution: Replace new cable. ►Problem & Solutoin Brand Olympus Model OTV-S6H-1D Fault Cable aging and interference Solution Replace cable ► Applications Laparoscopy (abdominal surgery) Arthroscopy (joint surgery) ENT (Ear, Nose, Throat) procedures General endoscopic imaging ► Troubleshooting & Support If the camera is not functioning, check: Cable connections
Quality S5-1 sector array ultrasound probe for repair factory

S5-1 sector array ultrasound probe for repair

S5-1 sector array ultrasound probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin ModelS5-1FaultArray element aging, lens bubbleSolutionReplace Crystal ► Common issues with S5-1 Probe Image Dropouts / Dead ElementsAcoustic Lens DamageCable DamageConnector DamageWeak or Noisy Image ►FAQ What causes "dark shadows" or vertical dropouts on the S5-1 screen, and can it be fixed? A: Vertical black lines or blind
Quality OLYMPUS CF-H180AL Electronic Colonoscope For Repair factory

OLYMPUS CF-H180AL Electronic Colonoscope For Repair

OLYMPUS CF-H180AL Electronic Colonoscope For Repair ► Brief Description Inspection Results: 1. The image is dark, and the CCD is aged. 2. One branch of the light beam is disconnected. 3. The insertion tube is aged and wrinkled. 4. The lower angle is malfunctioning, the other angles are tight, and the snake bone is deformed. 5. The light guide tube is slightly wrinkled. Solution: Repair and Replacement: CCD, light bundle, insertion tube, biopsy tube, snake bone, steel wire,
Quality OLYMPUS TJF-Q180V Electronic Duodenoscope For Repair factory

OLYMPUS TJF-Q180V Electronic Duodenoscope For Repair

OLYMPUS TJF-Q180V Electronic Duodenoscope For Repair ► Brief Description Inspection Results: 1. Image display is available; 2. The light bundle brightness is acceptable; 3. The insertion tube is worn; 4. The light guide tube is worn; 5. The angle is insufficient, the snake bone, steel wire and helix tube are worn; 6. The forceps elevator handle is damaged; 7. The handle shell is severely cracked. Solution: Replacement: Insertion tube, handle shell, light guide tube, snake
Quality Aloka UST-5413 linear array ultrasound probe for repair factory

Aloka UST-5413 linear array ultrasound probe for repair

Aloka UST-5413 linear array ultrasound probe for repair ► Brief Description Fault: Lens break. Solution: Replace lens. ►Problem & Solutoin BrandAlokaModelUST-5413FaultLens breakSolutionReplace lens ► Common issues with UST-5413 Probe Image dropouts or black linesAcoustic lens damageCable damageConnector damageWeak or noisy image ►FAQ How can we prevent the acoustic lens on the UST-5413 from peeling after a repair? Avoid Solvent Cleaners: Never use solutions containing high
Quality Ultrasonix EC9-5/10 Endo-cavitary probe for repair factory

Ultrasonix EC9-5/10 Endo-cavitary probe for repair

Ultrasonix EC9-5/10 Endo-cavitary probe for repair ► Brief Description Fault: Bad array element. Solution: Replace crystal and housing. ►Problem & Solutoin BrandUltrasonix ModelEC9-5/10FaultBad array elementSolutionReplace crystal and housing ► Common issues with EC9-5/10 Probe Image Dropouts or Streaks (Dead Elements)Acoustic Lens DamageCable and Strain Relief DamageHousing or Casing Damage ►FAQHow can I tell if my EC9-5/10 has a crystal failure or a cable failure?Crystal
Quality GE 4D3C-L 4D Convex Transducer For Repair factory

GE 4D3C-L 4D Convex Transducer For Repair

GE 4D3C-L 4D Convex Transducer For Repair ► Brief Description Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Solution:Replace the dome, dome tube; repair electric motor. ►Problem & Solutoin BrandGEModel4D3C-LFault1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged.SolutionReplace the dome, dome tube; repair electric motor ►Common issues with 4D3C-L 4D
Quality PENTAX Medical EPK-i7000 Video Processor For Repair factory

PENTAX Medical EPK-i7000 Video Processor For Repair

PENTAX Medical EPK-i7000 Video Processor For Repair ► Brief Description Brand: PENTAX (PENTAX Endoscope Processor) Model: EPK-i7000 Serial Number: AC*** Fault: It cannot recognize the endoscope and reports an error. Solution: Repair the system board. Repair Time: 3-5 working days. Warranty Period: Six months. ►Problem & Solution Brand PENTAX Model EPK-i7000 Product Processor Fault It cannot recognize the endoscope and reports an error. Solution Repair the system board. ►
Quality S5-2 Cardiac probe for repair factory

S5-2 Cardiac probe for repair

S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain relief ► Common issues with S5-2 Cardiac probe Poor or blurry images Constant noise or static Color streaking or intermittent artifacts Cracked or dried-out lens Cable or strain
Quality L15-7io Linear probe for repair factory

L15-7io Linear probe for repair

L15-7io Linear probe for repair ► Brief Description Fault: Lens problem. Solution: Replace probe lens. ►Problem & Solutoin Model L15-7io Fault Lens problem Solution Replace probe lens ► Common issues with L15-7io probe Cable and Connector Issues Acoustic Lens Damage Crystal/Element Failure Housing/Physical Damage Oil Leak If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.
Quality C5-1 convex probe for repair factory

C5-1 convex probe for repair

C5-1 convex probe for repair ► Brief Description Fault:Lens problem, cable broken, without strain relief. Solution:Replace probe lens, replace a new cable, replace strain relief. ►Problem & Solutoin Model C5-1 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, replace a new cable, replace strain relief ► Common issues with C5-1 probe Cable and Connector Damage Acoustic Lens Damage Crystal/Element Failure Intermittent Signal Loss or Noise
Quality EPlQ/X5-1 Probe 2D Image Abnormal For Repair factory

EPlQ/X5-1 Probe 2D Image Abnormal For Repair

EPlQ/X5-1 Probe For Repair ► Brief Description Fault: 2D image abnormal Solution: Replace scanhead ►Problem & Solutoin Model EPlQ/X5-1 Fault 2D image abnormal Solution Replace scanhead ► Common Problems & Symptoms Image Artifacts (black lines, dropouts, flickering) → Damaged elements or cable No Image/Dead Probe → Cable break, connector issues, or internal failure Lens Damage (cracks, delamination) → Acoustic coupling issues Error Messages ("Probe Not Recognized") → Faulty
Quality C5-2 Convex Probe For Repair Medical Device Parts factory

C5-2 Convex Probe For Repair Medical Device Parts

C5-2 Convex Probe For Repair ► Brief Description Fault: Strain relief damage, cable broken. Solution: Replace strain relief, repair cable. ►Problem & Solutoin Model C5-2 Convex Fault Strain relief damage, cable broken Solution Replace strain relief, repair cable ► Common Issues with C5-2 Probes Dead or flickering elements (image dropouts, black lines) Cable damage (broken wires near the connector or handle) Lens damage (cracks, delamination, or wear) Acoustic array failure
Quality L12-5 linear array ultrasound probe for repair factory

L12-5 linear array ultrasound probe for repair

L12-5 linear array ultrasound probe for repair ► Brief Description Fault: Crystal damage. Solution: Replace crystal. ►Problem & Solutoin Model L12-5 Fault Crystal damage Solution Replace crystal ► Common issues with L12-5 Probe Image Dropouts / Dead Elements Acoustic Lens Damage Cable Damage Connector Damage Noisy or Attenuated Image If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve
Quality L12-3 linear Probe for repair factory

L12-3 linear Probe for repair

L12-3 linear Probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin Model L12-3 Fault Array element aging, lens bubble Solution Replace Crystal ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief Separation Membrane/Lens Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will
Quality L12-3 Ultrasuound probe for repair factory

L12-3 Ultrasuound probe for repair

L12-3 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Model L12-3 Fault Lens problem Solution Replace lens ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief Separation Membrane/Lens Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.
Quality 1898200T M4 Straightshot Microdebrider and Handpiece for repair factory

1898200T M4 Straightshot Microdebrider and Handpiece for repair

1898200T M4 Straightshot Microdebrider and Handpiece for repair ► Brief Description Fault: 1) The bearing is corroded and the seal is broken. 2) Motor stuck, inside is full of disinfection residue, motor corrosion. 3) The rotation direction of the knife head is stuck. 4) Handle paint off, internal corrosion.. Solution: Exchange the whole handpiece except cable. ►Problem & Solutoin Model 1898200T Fault 1) The bearing is corroded and the seal is broken. 2) Motor stuck, inside
Quality Medical Device Repair Cases Stryker 30º 0502-503-030 Laproscope For Repair factory

Medical Device Repair Cases Stryker 30º 0502-503-030 Laproscope For Repair

Stryker 30º 0502-503-030 laproscope, SN:103**** Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. Repair cost: $*** Common Issues with the Stryker 0502-503-030 Laparoscope: 1.Fogging/Obscured Vision: This is a frequent issue during laparoscopic procedures. It can be caused by temperature differences between the endoscope and
Quality Medical Device Repair Cases LINVATEC HD 4300 Arthroscope For Repair factory

Medical Device Repair Cases LINVATEC HD 4300 Arthroscope For Repair

LINVATEC HD 4300 Arthroscope for repair, SN: 62**** Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Replace objective lens, inner optical parts, clean and adjust optical system, seal with eyepiece. Image Quality Related Issues: 1.Fogging or Blurry Vision: This is a frequent problem in arthroscopy. It can be due to condensation inside the endoscope, debris on the lens, or issues with the camera system. 2.Dark or Dim Image: This could be
Quality FUJINON EG-530N gastroscope for repair factory

FUJINON EG-530N gastroscope for repair

FUJINON EG-530N gastroscope for repair ► Brief Description Fault: 1.Biopsy tube broken; 2.Angle is not enough; 3.Camera over-heating, there are occasional snowflakes on the image and big risk for CCD; 4.Insert tube aging and wrinkle; 5.Light guide tube aging and wrinkle; 6.Stress boot of light guide tube broken; 7.Light bundle broken; 8.EVE tube aging and wrinkle. Solution: 1: Replace CCD; insert tube; light bundle; biopsy tube; bending mesh; wires; bending section; rubber;
Quality Olympus A4674A  Autoclavable Hysteroscope for repair factory

Olympus A4674A Autoclavable Hysteroscope for repair

Olympus A4674A Autoclavable Hysteroscope for repair ► Brief Description Fault: 1. Image blur and black shadow. 2. Rod lens broken. 3. Outer tube bending and transformation. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test. ►Problem & Solutoin Brand Olympus Model A4674A Fault 1. Image blur
Quality Karl Storz 27005BA Cystoscope for repair factory

Karl Storz 27005BA Cystoscope for repair

Karl Storz 27005BA Cystoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope and test. ►Problem & Solutoin