| Brand | Philips |
|---|---|
| Model | L15-7io |
| Fault | Lens Problem |
| Solution | Replace Probe Lens |
| Packaging Details | Standard Package |
| Brand | OLYMPUS |
|---|---|
| Model | TJF-Q180V |
| Fault | 1. Image Ok; 2. Light Bundlestill Ok; 3. Insert Tubestill Ok; 4. Biopsy Tubewear; 5. Bendingsection, Wires And Helix Tube Wear; 6. Knob Leakage. |
| Solution | Suggest Toreplace Biopsy Tube, Bending Section, Wires, Helix Tube, Knob And Rubber. (Nowarranty For CCD) |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | CF-240L |
| Fault | Leakage, Glass Of Light Bundle Blur, Rubber Leak, Angle Not Enough,image Fuzzle,objective Lens Scratch, |
| Solution | Replace Rubber, Objective Lens, C Cover. Adjust The Angle. |
| Packaging Details | Standard Package |
| Brand | Karl Storz |
|---|---|
| Model | 27005BA |
| Fault | 1. No Image. 2. Fiber Abrasion. 3. Rod Lens Broken And Cracked. 4. Objective Lens Broken And Dirty. 5. Eyepiece Cover Broken. |
| Solution | 1. Replace Rod Lens. 2. Replace Inner Tube. 3. Replace Outer Tube. 4. Replace Objective Lens Group. 5. Clean The Eyepiece. 6. Replace Fiber. 7. Replace Eyepiece Cover. 8. Clean And Install The Optical Path, Check The Image. 9. Seal With Endoscope A |
| Packaging Details | Standard Package |
| Brand | Toshiba |
|---|---|
| Model | PST-25BT |
| Fault | Housing Break, Image Weak, Crystal Aging |
| Solution | Replace Crystal And Housing |
| Packaging Details | Standard Package |
| Brand | STRYKER |
|---|---|
| Model | 1188-210-105 |
| Fault | Image With Interrupt. Cable Aging |
| Solution | Replace Compatible New Cable And Connector |
| Packaging Details | Standard Package |
| Brand | STORZ |
|---|---|
| Model | H3-ZA |
| Fault | Image Streaks Frequently,cable Loose Contact And Aged |
| Solution | Replace Cable Kit |
| Packaging Details | Standard Package |
| Brand | GE |
|---|---|
| Model | RIC6-12-D |
| Fault | Lens Problem |
| Solution | Replace Lens,Clean Cable |
| Packaging Details | Standard Package |
| Brand | Hitachi |
|---|---|
| Model | C41V1 |
| Fault | Lens Problem |
| Solution | Replace Lens |
| Packaging Details | Standard Package |
| Brand | Medtronic |
|---|---|
| Model | 1898200T |
| Fault | 1) The Bearing Is Corroded And The Seal Is Broken. 2) Motor Stuck, Inside Is Full Of Disinfection Residue, Motor Corrosion. 3) The Rotation Direction Of The Knife Head Is Stuck. 4) Handle Paint Off, Internal Corrosion. |
| Solution | Exchange The Whole Handpiece Except Cable. |
| Packaging Details | Standard Package |