Medical Device Repair Cases

Quality LINVATEC HD 4300 Arthroscope for repair factory

LINVATEC HD 4300 Arthroscope for repair

LINVATEC HD 4300 Arthroscope for repair ► Brief Description Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Solution: Replace objective lens, inner optical parts, clean and adjust optical system. ►Problem & Solutoin Brand LINVATEC Model HD 4300 Fault Image with black shadow; eyepiece with scratches, inner optical parts broken. Solution Replace objective lens, inner optical parts, clean and adjust optical system. ► Common issues with HD
Quality Stryker 30º 0502-503-030 laproscope for repair factory

Stryker 30º 0502-503-030 laproscope for repair

Stryker 30º 0502-503-030 laproscope for repair ► Brief Description Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. ►Problem & Solutoin Brand Stryker Model 0502-503-030 Fault Image blur, objective lens loose without glue, inner optical parts broken. Solution Add glue for objective lens, replace inner optical parts, clean and
Quality Olympus A70940A  Autoclavable Arthroscope for repair factory

Olympus A70940A Autoclavable Arthroscope for repair

Olympus A70940A Autoclavable Arthroscope for repair ► Brief Description Fault: 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test. ►Problem & Solutoin Brand Olympus Model
Quality Olympus GIF-Q150 Gastroscope for repair factory

Olympus GIF-Q150 Gastroscope for repair

Olympus GIF-Q150 Gastroscope for repair ► Brief Description Fault: 1. With image; 2. Light bundlebreak badly; 3. Insert tubeaging; 4. Biopsy tubeabrasion, leakage badly; 5. Angle tied, bendingsection, wires and helix tube abrasion; 6. Rubberwithout glue. Solution: Replace light bundle, insert tube, biopsy tube, bending section, wires, helixtube and rubber. (No warranty for CCD) ►Problem & Solutoin Brand Olympus Model GIF-Q150 Fault 1. With image; 2. Light bundlebreak badly; 3
Quality GIF-XQ240 Gastroscope by Olympus for repair factory

GIF-XQ240 Gastroscope by Olympus for repair

GIF-XQ240 Gastroscope by Olympus for repair ► Brief Description Fault: Leakage. Solution: Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ►Problem & Solutoin Brand Olympus Model GIF-XQ240 Fault Leakage. Solution Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ► Common issues with GIF-XQ240 Gastroscope No Image / Dark Image Poor Image Resolution / Blurry Image Ineffective Suction Difficulty with Angulation (Stiff or Loose) Bubbles in the Field of View
Quality Olympus CYF V2 Cystoscope For Repair factory

Olympus CYF V2 Cystoscope For Repair

Olympus CYF V2 Cystoscope For Repair ► Brief Description Fault: Leakage. Solution: Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber. ►Problem & Solutoin Brand Olympus Model CYF V2 Fault Leakage Solution Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber ► Common issues with CYF V2 Cystoscope No Image / Dark Image Poor Image Resolution / Blurry Image Ineffective Suction Difficulty
Quality Olympus GIF-H170 gastroscope for repair factory

Olympus GIF-H170 gastroscope for repair

Olympus GIF-H170 gastroscope for repair ► Brief Description Fault: 1. The top left and middle right side with blur image, objective Lens broken. 2. 20% light bundle broken, the small lens turned yellow. 3. Insert tube yellow and wrinkles. 4. The biopsy tube abrasion. 5. The wire and helix tube abrasion. 6. Light guide tube aging and wrinkle. Solution: Replace objective lens, light bundle, small lens, insert tube, biopsy tube, wires, helix tube, light guide tube and rubber.
Quality R Wolf 7326071 ureteroscope for repair factory

R Wolf 7326071 ureteroscope for repair

R Wolf 7326071 ureteroscope for repair ► Brief Description Fault: The image has spots,image bundle aging,light bundle broken;Insert tube aging;Biopsy channel worn and leakage;Bending section,wire and helix tube were worn. Solution: Replace the light bundle,image bundle,rubber,biopsy channel,helix tube,wire. ►Problem & Solutoin Brand R Wolf Model 7326071 Fault The image has spots,image bundle aging,light bundle broken;Insert tube aging; Biopsy channel worn and leakage;Bending
Quality Insighters IS3-F laryngoscope for repair factory

Insighters IS3-F laryngoscope for repair

Insighters IS3-F laryngoscope for repair ► Brief Description Fault: 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution: Replace CCD, insert tube, bending section, rubber ►Problem & Solutoin Brand Insighters Model IS3-F Fault 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution Replace CCD, insert tube, bending section, rubber ► Common
Quality KARL STORZ 11278A1 ureteroscope for repair factory

KARL STORZ 11278A1 ureteroscope for repair

KARL STORZ 11278A1 ureteroscope for repair ► Brief Description Fault: 1) Image spotted with partially black point, image bundle aged 2) Light bundle partially wire broken 3) Insert tube aged 4) Biopsy tube abrasion and leakage 5) Wires and coil pipe abrasion Solution: Replace HD image bundle, light bundle, insert tube, wires, coil pipe and rubber. ►Problem & Solutoin Brand KARL STORZ Model 11278A1 Fault 1) Image spotted with partially black point, image bundle aged 2) Light
Quality Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair factory

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair ► Brief Description Fault: 1. Image with black point and image bundle broken 2. Light bundle aged 3. Insert tube aging and wrinkle 4. Bending section deformation 5. Angle with not enough 6. Rubber abrasion Solution: Replace image bundle; light bundle; insert tube; wires; helix tube; bending section; rubber ►Problem & Solutoin Brand Olympus Model ENF-GP Fault 1. Image with black point and image bundle broken 2. Light
Quality Olympus BF-XP160F Bronchoscope for repair factory

Olympus BF-XP160F Bronchoscope for repair

Olympus BF-XP160F fiber scope for repair ► Brief Description Fault: 1. Image degumming, serious color, CCD aging, a few black and gray spots; 2. 20% of the light bundle is broken; 3. Insert tube aging slightly; 4. Biopsy tube wear, not smooth; 5. Bending section, steel wire, screw tube wear; Solution: Replace CCD, image beam, beam, insert tube, pliers tube, snake bone, steel wire, spiral tube, rubber. ►Problem & Solutoin Brand Olympus Model BF-XP160F Fault 1. Image degumming,
Quality Olympus CF-2T160L Colonoscope for repair factory

Olympus CF-2T160L Colonoscope for repair

Olympus CF-2T160L Colonoscope for repair ► Brief Description Fault: 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy tube abrasion. 5. Angle tied, bending section transformation. 6. Wires, helix tube abrasion, no up angle. Solution: Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber ►Problem & Solutoin Brand Olympus Model CF-2T160L Fault 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy
Quality Olympus CF-Q160AL Colonoscope for repair factory

Olympus CF-Q160AL Colonoscope for repair

Olympus CF-Q160AL scope for repair ► Brief Description Fault: 1.No image, CCD broken; 2.image broken completely; 3.insert tube aging; 4.biopsy tube abrasion; 5.bending section abrasion; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace CCD, light bundle, insert tube,biopsy tube,aire&water tube,bending section,wires,helix tube,light guide tube,all the fstener part,clamp and rubber. ►Problem & Solutoin Brand Olympus Model CF-Q160AL Fault 1
Quality Olympus CHF-P20Q endoscope for repair factory

Olympus CHF-P20Q endoscope for repair

Olympus CHF-P20Q endoscope for repair ► Brief Description Fault: 1.Insert tube cracked; 2.Rubber leakage; 3.Image with black point and cracked; 4.Angulation tight; 5.Biopsy tube aging Solution: Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber ►Problem & Solutoin Brand Olympus Model CHF-P20Q Fault 1.Insert tube cracked; 2.Rubber leakage; 3.Image with black point and cracked; 4.Angulation tight; 5.Biopsy tube aging Solution Replace image
Quality Olympus GIF-1TQ160 gastroscope for repair factory

Olympus GIF-1TQ160 gastroscope for repair

Olympus GIF-1TQ160 gastroscope for repair ► Brief Description Fault: 1.With image, but CCD with broken. 2.Light bundle broken completely. 3.Insert tube abrasion. 4. No biopsy tube. 5. No air&water tube, no water supply tube. 6. No bending section assembly. 7. No digital tip. Solution: Replace CCD, digital tip, light bundle, insert tube,biopsy tube, air&water tube, water supply tube,bending section,wires,helix tube and rubber. ►Problem & Solutoin Brand Olympus Model GIF-1TQ160
Quality Olympus GIF-H180J gastroscope for repair factory

Olympus GIF-H180J gastroscope for repair

Olympus GIF-H180J gastroscope for repair ► Brief Description Fault: 1.image with shadow; 2.light bundle broken; 3.insert tube with bite mark; 4.biopsy tube abrasion; 5.bending section serious deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace CCD, light bundle, insert tube,biopsy tube,aire&water tube,bending section,wires,helix tube,light guide tube and rubber. ►Problem & Solutoin Brand Olympus Model GIF-H180J Fault 1.image
Quality Olympus GIF-Q180 gastroscope for repair factory

Olympus GIF-Q180 gastroscope for repair

Olympus GIF-Q180 gastroscope for repair ► Brief Description Fault: 1.image definition not enough, CCD aging. 2.30% light bundle broken. 3. Insert tube aging and wrinkle. 4.Biopsy tube abrasion. 5. Bending section bited, transformation. 6. Angle not enough, wires and helix tube abrasion. 7. Light guide tube aging and wrinkle. 8. Pins for wire holder leakage. Solution: Replace CCD, light bundle,insert tube, biopsy tube, bending section,helix tube, wires, light guide tube, pins
Quality Olympus CF-240AI Colonoscope for repair factory

Olympus CF-240AI Colonoscope for repair

Olympus CF-240AI Colonoscope for repair ► Brief Description Fault: Image blur,insert tube aging,light guide squashed,light guide broken 50%,nozzle blocked,angle tight,forceps channel aging. Solution: Replace insertion tube,CCD chips,light guide,forceps channel,water channel,angulation wire,bending mesh,spiral tube,light guide tube,bending rubber. ►Problem & Solutoin Brand Olympus Model CF-240AI Fault Image blur,insert tube aging,light guide squashed,light guide broken 50%
Quality Olympus LTF-160 thoracoscope for repair factory

Olympus LTF-160 thoracoscope for repair

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the buttons failure. 8. Rubber leakage. 9. the whole scope leakage. Solution: Replace CCD,light bundle, biopsy tube, wires,helix tube,grip and rubber, repair all the buttons. ►Problem &
Quality Olympus TJF-160VR endoscope for repair factory

Olympus TJF-160VR endoscope for repair

Olympus TJF-160VR endoscope for repair ► Brief Description Fault: 1.No image, CCD broken. 2. Light bundle broken completely. 3. Insert tube aging and wrinkle. 4. Biopsy tube broken. 5. Wires and helix tube broken. 6. clamp wires broken. 7. Operation part broken, no angle. 8. Grip broken. 9. The scope be separated by someone before, some parts missing. Solution: Replace digital parts,CCD,light bundle,insert tube, biopsy tube,air&water tube, bending section, wires,helix tube,
Quality Olympus GIF-Q150 gastroscope for repair factory

Olympus GIF-Q150 gastroscope for repair

Olympus GIF-Q150 gastroscope for repair ► Brief Description Fault: 1.with image; 2.one eye of light bundle broken; 3.insert tube little aging; 4.biopsy tube abrasion; 5.bending section deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace light bundle ,insert tube, biopsy tube,bending section,wires,helix tube,light guide tube and rubber. ►Problem & Solutoin Brand OLYMPUS Model GIF-Q150 Fault 1.with image; 2.one eye of light
Quality STORZ 13820pks Gastroscope for Repair factory

STORZ 13820pks Gastroscope for Repair

STORZ 13820pks Gastroscope for Repair ► Brief Description Fault: 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of. Solution: 1. Replace the wires, coil pipe, rubber; 2.Adjust angle. ►Problem & Solutoin Brand STORZ Model 13820pks Fault 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of Solution 1. Replace the wires, coil pipe, rubber; 2.Adjust angle ► Common issues with 13820pks Gastroscope Dim or Uneven Illumination Fuzzy, Blurry, or
Quality Olympus CF-H190L Colonoscope for repair factory

Olympus CF-H190L Colonoscope for repair

Olympus CF-H190L Colonoscope for repair ► Brief Description Fault: 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; 6. Soft and hard adjustment wire wear, relatively tight; 7. Rubber leakage. Solution: Replace the insertion tube, biopsy tube, bending section, wire, helix tube, soft and hard