Quality Medical Device Repair Cases OF GE RIC5-9-D ULTRASOUND PROBE FOR REPAIR factory

Medical Device Repair Cases OF GE RIC5-9-D ULTRASOUND PROBE FOR REPAIR

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...

Quality OLYMPUS CF-HQ190L Medical Device Repair Cases Repair Tool Case factory

OLYMPUS CF-HQ190L Medical Device Repair Cases Repair Tool Case

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...

Quality GE 4D3C-L 4D Convex Transducer For Repair factory

GE 4D3C-L 4D Convex Transducer For Repair

GE 4D3C-L 4D Convex Transducer For Repair ► Brief Description Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Solution:Replace the dome, dome tube; repair electric motor. ►Problem & Solutoin BrandGEModel4D3C-LFault1.Probe dome ...

Quality GE 4C-RS abdominal probe for repair factory

GE 4C-RS abdominal probe for repair

GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...

Quality S5-2 Cardiac probe for repair factory

S5-2 Cardiac probe for repair

S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain ...

Quality Olympus LTF-160 thoracoscope for repair factory

Olympus LTF-160 thoracoscope for repair

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the ...

Quality OLYMPUS CF-H180AL Electronic Colonoscope For Repair factory

OLYMPUS CF-H180AL Electronic Colonoscope For Repair

OLYMPUS CF-H180AL Electronic Colonoscope For Repair ► Brief Description Inspection Results: 1. The image is dark, and the CCD is aged. 2. One branch of the light beam is disconnected. 3. The insertion tube is aged and wrinkled. 4. The lower angle is malfunctioning, the other angles are tight, and ...

Quality SonoScape EG-500 Electronic Gastroscopy For Repair factory

SonoScape EG-500 Electronic Gastroscopy For Repair

SonoScape EG-500 Electronic Gastroscopy For Repair ► Brief Description Inspection Results: 1. No image display, CMOS water ingress fault; 2. 20% of the light bundle is broken; 3. There are bite marks on the insertion tube in multiple places; 4. The biopsy tube is worn out and not smooth; 5. The ...

Quality Olympus Flexible Endoscope Air & Water Tube A/W Transparent Tube factory

Olympus Flexible Endoscope Air & Water Tube A/W Transparent Tube

Olympus/ Pentax/ Fuji A/W Tube ► Brief Description The "air/water channel" is a critical part of a flexible endoscope's function. The tube you are referring to is most commonly known as the Air/Water Valve Tube or A/W Tube . It's a small, transparent, and highly flexible tube that connects the air ...