| Brand | Olympus |
|---|---|
| Model | CF-240L |
| Fault | Leakage, Glass Of Light Bundle Blur, Rubber Leak, Angle Not Enough,image Fuzzle,objective Lens Scratch, |
| Solution | Replace Rubber, Objective Lens, C Cover. Adjust The Angle. |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | CF-H190L |
| Fault | 1. Image Ok; 2. Light Bundle Still Ok; 3. The Front End Of The Insert Tube Is Broken, And The Serpentine; The Position Of Stress Boot Is Leaking; 4. Biopsy Tube Wear; 5. Left Angle Failure, Bending Section, Wire And Helix Tube Wear; 6. Soft And Hard |
| Solution | Replace The Insertion Tube, Biopsy Tube, Bending Section, Wire, Helix Tube, Soft And Hard Adjusting Wire And Rubber. (No Warranty For CCD) |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | GIF-Q180 |
| Fault | 1.image Definition Not Enough, CCD Aging. 2.30% Light Bundle Broken. 3. Insert Tube Aging And Wrinkle. 4.Biopsy Tube Abrasion. 5. Bending Section Bited, Transformation. 6. Angle Not Enough, Wires And Helix Tube Abrasion. 7. Light Guide Tube Aging And |
| Solution | Replace CCD, Light Bundle,insert Tube, Biopsy Tube, Bending Section,helix Tube, Wires, Light Guide Tube, Pins And Rubber. |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | GIF-1TQ160 |
| Fault | 1.With Image, But CCD With Broken. 2.Light Bundle Broken Completely. 3.Insert Tube Abrasion. 4. No Biopsy Tube. 5. No Air&water Tube, No Water Supply Tube. 6. No Bending Section Assembly. 7. No Digital Tip. |
| Solution | Replace CCD, Digital Tip, Light Bundle, Insert Tube,biopsy Tube, Air&water Tube, Water Supply Tube,bending Section,wires,helix Tube And Rubber. |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | CF-Q160AL |
| Fault | 1.No Image, CCD Broken; 2.image Broken Completely; 3.insert Tube Aging; 4.biopsy Tube Abrasion; 5.bending Section Abrasion; 6.wires And Helix Tube Abrasion; 7.light Guide Tube Aging And Wrinkle. |
| Solution | Replace CCD, Light Bundle, Insert Tube,biopsy Tube,aire&water Tube,bending Section,wires,helix Tube,light Guide Tube,all The Fstener Part,clamp And Rubber. |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | CF-2T160L |
| Fault | 1. With Image. 2. 10% Light Bundle Broken. 3. Insert Tube Is Ok. 4. Biopsy Tube Abrasion. 5. Angle Tied, Bending Section Transformation. 6. Wires, Helix Tube Abrasion, No Up Angle. |
| Solution | Replace Wires, Helix Tube, Bending Section And Rubber. |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | CF-240AI |
| Fault | Image Blur,insert Tube Aging,light Guide Squashed,light Guide Broken 50%,nozzle Blocked,angle Tight,forceps Channel Aging. |
| Solution | Replace Insertion Tube,CCD Chips,light Guide,forceps Channel,water Channel,angulation Wire,bending Mesh,spiral Tube,light Guide Tube,bending Rubber. |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | CYF V2 |
| Fault | Leakage |
| Solution | Replace The Insert Tube, Light Bundle, Biopsy Channel, Bending Section, Wire, Steel Mesh, Rubber |
| Packaging Details | Standard Package |
| Brand | Olympus |
|---|---|
| Model | GIF-XQ240 |
| Fault | Leakage. |
| Solution | Replace Light Guide Tube,NO.1 Button,wire,mesh,adjust Angle. |
| Packaging Details | Standard Package |